New Analytical Formulae to Calibrate HPGe Well-type Detectors Efficiency and to Calculate Summing Corrections

Abstract

Abbas MI and Ibrahim OA

Direct mathematical method is used in this paper to calculate the photo peak efficiency of a system containing a radioactive source placed in the well of a well type detector. Attenuation of the source and detector’s walls are included. In addition summing corrections are also included. The comparison between published experimental values and present calculated values are given. This work proved quite success in predicting the photo peak efficiency of certain detectors even for a radionuclide emitting rays with different energies.